ULTRASOUND AS A WAVE
Sound is propagated
through a medium (e.g. air) as a mechanical vibration of the particles
of that medium and in simple terms may be categorised by its loudness
and pitch or frequency. “Ultra” means beyond, ultrasound is sound with a
frequency beyond that of human perception (i.e. >20 kHz), and
has the same physical properties as “audio” sound. Most clinical
diagnostic applications of ultrasound employ frequencies in the range 2 -
10 MHz.
Ultrasonic energy travels through a medium in
the form of a wave. Although a number of different wave modes are
possible, in almost all diagnostic applications, ultrasound propagates
in the form of a longitudinal wave, where the particles of the medium
oscillate in the direction of propagation of the sound. Energy is
transferred through the medium in a direction parallel to that of the
oscillations of the particles. The particles themselves do not move
through the medium. They simply vibrate to and fro about their mean
position.
The vibrations of individual particles may be
complex For simplicity consider the movement of a single particle
excited by pure sinusoidal continuous wave.
The graph
above shows the displacement of the particle about its mean position
plotted against time. The time taken to execute one complete cycle, T,
is called the period. The maximum displacement, a, is known as the
amplitude. If the frequency of the wave is f (Hz) i.e. it executes f
complete cycles per second. The time taken to execute one complete
cycle, T, is given by
Period = 1/f seconds
It
is often useful to think of the source of ultrasound, the transducer,
as a vibrating piston. As it moves it displaces the adjacent particles
of the adjoining medium. These in turn displace more particles
throughout the medium. Since the particles are not rigidly fixed to each
other, they do not all move together. There is a delay between the
movement of adjacent particles (analogous to a series of balls connected
by springs). At a particular time there will be some regions where the
particles are closer together and the pressure and density of the medium
is increased (regions of compression) and areas where the particles are
further apart and the pressure and density of the medium is decreased
(regions of ). These regions of compression or rarefaction move through
the medium as a wave.
compression
rarefraction
Consider again the case of
excitation by a simple sinusoidal waveform. At a given time, the
displacement of the particles (or the pressure or density of the
particles) plotted against distance is shown below.
The
wavelength, lambda, is defined as the distance in the medium between
points of equal value (displacement, pressure or density). For a sound
wave of wavelength and frequency f, the distance travelled by the
wave per second (i.e. its velocity, c) will be the number of cycles
passing a given point in unit time multiplied by the length of each
wave.
i.e. c = f.lambda
The velocity
of a sound wave in a medium is determined by the delay which occurs
between the movements of neighbouring particles. This depends on the
elasticity and density of the medium. It is a property of the medium and
is essentially independent of f, the frequency of the ultrasound.
Rigid materials have higher wave velocities than compressible materials
like gases.
MATERIAL Velocity c (m s-1)
Air
330
Lung 650
Water 1480
Aluminium 6400
Bone
3500
Brain 1540
Blood 1570
Fat 1460
Muscle
1580
Soft tissue (average) 1540
The Table above
shows that the values for the speed of sound in different tissues are
very similar. It is assumed that a value for wave velocity of 1540 m/s
is a reasonable approximation in most clinical applications. This is of
fundamental importance in diagnostic use of ultrasound.
INTENSITY
When
an ultrasound wave is generated in tissue, energy, in the form of
kinetic energy of motion of the particles, passes through the tissue.
The intensity, I, of an ultrasound field is the quantity of energy
flowing through unit area in unit time. i.e.
where E is the total
energy in J, A is the area, t = time
Intensity is usually
measured in units of Webers per square meter (W m-2) or mW cm-2
Intensity
is not normally measured directly, calibrated transducers known as
hydrophones are usually used to measure pressure amplitudes within
ultrasound beams. It can be shown that
where P is the
pressure amplitude and is the density
Intensity is
proportional to the square of the pressure.
For
continuous wave fields, provided the measurement is carried out over
many cycles, the intensity will always have the same value. However for
pulsed fields, intensity figures are usually quoted in terms of temporal
average, pulse average or temporal peak. The temporal average intensity
ITA is the measurement obtained after averaging over many cycles. The
pulse average intensity IPA is the value obtained by averaging only
during the duration of a pulse and not during the 'off-time'. The
temporal peak intensity lTP is the maximum instantaneous value
measurable and corresponds to the peak value of the pulse.
Spacial
variation must be considered. Within the beam there will be areas of
high intensity and areas of low intensity. The regions of maximum
intensity might for example be at the focus of a focused system. This
maximum may extend for only a few mm in any direction. If the intensity
is measured over that small region, the value will be high. On the other
hand if it is averaged over the whole cross-section a much smaller
value will result. It has become conventional to refer to spatial peak
and spatial average intensities. The spatial peak intensity ISP value is
the maximum intensity sampled over a very small distance found anywhere
in the beam. The spatial average intensity ISA value is the average
value across the beam at some distance from the transducer.
Combining
the above concepts we get a whole range of intensity parameters. It is
ISPTA (Spatial Peak Temporal Average) value which is most critical as it
relates to the local heating effect in tissue.
DECIBEL
NOTATION
An absolute measurement of intensity is
difficult and often inappropriate. Usually, we are more interested in
knowing the ratio of intensities. particularly if the level of one of
these is taken as a reference for comparison (e.g. ratio of energy
reflected at a different tissue boundaries). Expressing such ratios as
logarithms provides a simple method of expressing numbers which extend
over many orders of magnitude.
The relative intensity in decibels
(dB) where I1 and I0 are the intensities
or
The
relative amplitude in decibels (dB) where A1 and A0 are the wave
amplitudes
Where I1 > I0, dB values are +ve, where I1
< I0, dB values are -ve.
DECIBEL
NOTATION FOR INTENSITIES
dB I1/I0 dB I1/I0
0 1 0 1
+3
1.995 -3 0.501
+10 10 (101) -10 0.1 (10-1)
+20 100 (102)
-20 0.01 (10-2)
+30 103 -30 10-3
+40 104 -40 10-4
DECIBEL
NOTATION FOR AMPLITUDES
dB A1/A0 dB A1/A0
0 1 0 1
+3
1.413 -3 0.708
+6 1.995 -6 0.501
+10 3.162 -10 0.320
+20
10 (101) -20 0.1 (10-1)
+40 102 -40 10-2
+60 103 -60
10-3
The existence of these two separate expressions
can lead to confusion. The first term is used when comparing intensity
or power, the second term is used when comparing pressure amplitude or
voltage.
In ultrasound systems, the intensity or power
of an ultrasound transducer is changed by varying the excitation
voltage. Since power is proportional to the voltage squared, V2, 10
log10 V2 = 20 log10 V. Similarly, intensity is measured by hydrophones
which measure pressure amplitude and intensity is proportional to the
pressure squared. Provided the correct definition is used there is a
complete equivalence between the decibel relative level for both terms.
For example, if the voltage gain of an output amplifier driving an
ultrasound transducer is increased by +3dB, the intensity or power of
the ultrasound wave will also increase by +3dB.
INTERFERENCE
When
two (or more) waves are transmitted into a medium the resultant
particle motion is obtained by adding the motion due to one wave to the
motion due to the other. This phenomenon is known as interference.
When
two ultrasound pressure waves with the same frequency and in step (in
phase) overlap (as in (a)), they reinforce each other and the resulting
waveform has an increased amplitude. The interference has resulted in a
wave with an increased intensity - this process is known as constructive
interference.
Conversely, when two waves with the same
frequency but out of step (1800 out of phase) combine (as in (b)), the
resultant amplitude would be small since the summation of the wave
motion would tend to cancel each other out. This is know as destructive
interference.
For interference to occur the waves have
to be coherent. The phase relationship must hold over many cycles and
the frequencies must be equal, or very nearly so.
PROPAGATION
OF ULTRASOUND THROUGH TISSUE
As an ultrasound wave
propagates through tissue, its intensity is attenuated by a number of
mechanisms. The ultrasound beam will diverge due to the difficulty of
generating a parallel beam (see later) and the refraction, reflection
and scattering of the ultrasound wave. Furthermore, the mechanical
energy of the ultrasound beam will be converted to heat by absorption.
As a rough rule of thumb, the total attenuation of soft tissue is
approximately 1 dB cm-1 MHz-1.
Refraction
When
a wave meets a boundary between two media at normal incidence (90º), it
is propagated without deviation into the second medium. At oblique
incidence the wave is bent by refraction. The amount determined by
Snell's law:
where c1 = wave velocity in
medium 1 and c2 = wave velocity in medium 2.
Ultrasound
refraction is normally insignificant in most areas of medical
ultrasound apart from the eye at the interface between aqueous and
vitreous humour or if trying to scan through bone. It can b significant
when using a test phantom.
Reflection
When
an ultrasound wave meets a boundary between two different media, where
the size of the boundary is large compared with the wavelength of
ultrasound and the roughness of the boundary is small compared with the
wavelength, a proportion of the ultrasound energy is reflected. This
specular reflection is similar to optical reflection i.e. *i = *r. In
normal incidence the reflected beam will return to the transducer along
the same path. This returned “echo” forms the basis of pulse echo
ultrasound imaging.
The proportion of the
incident energy reflected by the boundary is also important and depends
on the acoustic impedance (Z) of each medium.
Z * *c where * is
the density of the material.
In normal incidence the
fraction of the wave reflected is given by
where
Ir
= intensity of reflected ultrasound
Ii = intensity of incident
ultrasound
Z1¬ = acoustic impedance in medium 1
Z2 =
acoustic impedance in medium 2
Hence, it is
the difference between the acoustic impedance of the two structures that
determines the proportion of the incident energy that is reflected.
Examples of values of Z are given below together with examples of the
proportion of energy reflected at typical boundaries
MATERIAL
Z (106 kg m-2 s-1)
AIR 0.0004
LUNG 0.26 -
0.46
BONE 3.75 - 7.38
WATER 1.52
LIVER 1.65
KIDNEY
1.62
BLOOD 1.61
FAT 1.38
TISSUE 1.35 - 1.68
BOUNDARY
% REFLECTED
at normal incidence
MUSCLE/BLOOD
0.07%
FAT/MUSCLE 1.08%
SOFT TISSUE/WATER 0.23%
SOFT
TISSUE/AIR 99.9%
SOFT TISSUE/BONE 41.2%
When
Z1 = Z2, all the energy is transmitted across the boundary and there is
no reflected echo. However, when the difference in acoustic impedance
between medium 1 and 2 is large, (e.g. between a tissue/air or
tissue/bone interface) most of the ultrasound energy is reflected and
very little is transmitted. Hence it is difficult to visualise through
bone or through air in the lungs and bowel. In order to exclude air
between the transducer and skin surface a coupling gel is used to ensure
adequate penetration of ultrasound into the tissues.
Scattering
When
an ultrasound wave strikes targets which are small or rough compared to
the wavelength (e.g. within soft tissues, organs and blood), these
targets re-radiate (scatter) the ultrasound energy in many directions.
Where there are many scattering targets, multiple scattering occurs.
These
scatterers act as secondary sources of ultrasound.
A proportion
of this scattered ultrasound energy will return in the direction of the
source (back-scattered). The contribution which scattering makes to the
total attenuation is frequency dependent. When scatterer size
<< *, scattered intensity * f4. However the scattered
intensity is small in comparison to the energy reflected from major
tissue boundaries. For example, compared to the reflected intensity from
a fat/muscle boundary, the approximate scattered intensities from a
number of different structure are given below:
Placenta
-20dB (10-2)
Liver -30dB (10-3)
Kidney -40dB (10-4)
Blood
-60dB (10-6)
Between the strong directional echoes
from specular reflection at boundaries and the weak multi-directional
echoes from scattering targets, a range of intermediate echoes are
received. Any ultrasound system thus has to be capable of processing a
wide dynamic range of echoes.
Absorption
Absorption
is the process by which some of the mechanical energy of the ultrasound
is converted into heat in the tissues. In soft tissue, absorption
account for over 90% of the total attenuation of the ultrasound beam.
Absorption falls off exponentially with distance, the same fraction of
the incoming energy is lost in each unit distance travelled.
The
intensity, I, at a distance is given by:
where is
the initial intensity at x = 0,
and is the intensity absorption
coefficient.
The absorption coefficient depends on the
characteristics of the medium and is also approximately proportional to
ultrasound frequency. Hence in order to achieve greater penetration
(less attenuation) lower frequencies are used.
Instead
of giving values of , it more convenient to define the half-value
thickness as the thickness of material required to reduce the intensity
of an ultrasound beam by half, or -3dB.
HALF VALUE
THICKNESS (cm)
MATERIAL 2MHz 5MHz
Air
0.06 0.01
Bone 0.1 0.04
Water 340 54
Soft Tissue
2.1 0.86
Blood 8.5 3.0
Liver 1.5 0.5
Air
and bone have high values of . As well as strongly reflecting
ultrasound at any interface with tissue as mentioned earlier, they
attenuate the small proportion transmitted. Problems of scanning through
the head or imaging through the lungs or bowel are compounded. Water
and other body fluids have low attenuation, having a full bladder is a
standard technique to get good views of the uterus.
GENERATION
AND DETECTION OF ULTRASOUND
Ultrasound
is generated and detected by a transducer, which converts electrical
energy into mechanical vibrations and vice versa. Materials which
generate a potential difference across their surface when their shape is
changed by an applied pressure wave by the piezoelectric effect, are
used as transducers. These materials also change their shape when a
voltage is applied (the inverse piezoelectric effect) and are therefore
used as both transmitters and receivers of ultrasound. There are many
naturally occurring piezoelectric materials such as quartz, but it is
normal to use synthetic materials such as a ceramic - lead
zirconate/titanate (PZT) or a plastic - polyvinyldifluoride (PVDF)
transducer.
If a piezoelectric transducer is excited
with a continuous sinusoidal electrical signal, it will oscillate and
generate an ultrasonic wave at the same frequency as the excitation
frequency. Transducers display a natural frequncy where resonance occurs
and generation of ultrasound waves is particularly efficient.
When
a sinusoidal electrical signal is applied to a piezoelectric material
the walls will vibrate. Some of the enegry will travels into any
adjoining medium. A wave is also reflected inside the transducer and be
reflected to and fro. If the time taken for this internal wave to travel
from one side to the other and then back again is the same as the
period of the applied signal, the internally reflected waves interfere
constructively, and the resultant ultrasound wave is enhanced. This
first resonance occurs where the thickness of the material, t = /2,
therefore
the resonant frequency
Often only this
lowest resonant frequencies carries significant energy. If a transducer
is designed so that its thickness is equal to half the wavelength
corresponding to the required frequency of operation the transducer is
operating at its fundamental resonant frequency giving maximum
efficiency in transmission and reception.
If
the applied frequency is varied, the displacement of the transducer,
and hence the energy of the resulting ultrasound wave will decrease as
shown. The width of the peak, delta f, where the amplitude has fallen by
-3dB (A1/A0 delta 0.7) is known as the transducer bandwidth. The
bandwidth of commercial transducers may extend over several MHz. The
Q-factor, Q, of a transducer is defined as
Q = fr
/delta f .
Transducer design depends on the
mode of operation. In a simple transducer each face of the
piezoelectric element is coated with a thin metallic layer to act as an
electrode, and the complete assembly is housed in a metal cylinder. In
continuous wave applications and in order to obtain high efficiency, the
rear face of the piezoelectric material is backed by air. This helps to
reflect energy from the rear face back into the material to reinforce
the wave from the front face as described above. However most
applications use pulsed excitation in order to obtain a short burst or
pulse of ultrasound. In early instruments a voltage pulse was applied to
the transducer forcing it to oscillate at its natural resonant
frequency. Modern pulsed systems apply a few cycles of a sinusoidal
waveform. In both cases it is required to produce a short pulse of
ultrasound. Highly efficient air backed transducers are unsuitable since
the internally reflected signal would continue to produce ultrasound
after the applied electrical signal had stopped. This ringing can be
reduced by a backing material which reduces reflections at the back
face. Short pulse outputs of ultrasound can be achieved at the expense
of a less efficient, higher bandwidth (lower Q) transducer.
ULTRASONIC
FIELD
The ultrasonic field of a transducer
describes the spatial distribution of its radiated energy. The field
during transmission is identical to the sensitivity distribution of the
transducer when used as a receiver.
The ultrasound beam
shape produced by transducer is complex. In an idealised situation of a
circular transducer generating a continuous wave of ultrasound, the
beam shape can be considered to have two distinct regions.
In
the near field (or Fresnel region) the ultrasound beam is approximately
cylindrical with a diameter roughly equal to the transducer diameter.
The near field extends for a distance of D2/4* from the transducer face,
where D is the transducer diameter and * is the wavelength of the
ultrasound. In the far field (or Fraunhofer region) the beam diverges
with an angle given by sin* = 1.22*/D.
However, within
these two regions the beam intensity is not uniform and becomes even
more complex when rectangular, focused and pulsed transducers are used.
A
more accurate estimation of the field of a transducer can be obtained
by considering the surface of the transducer to be an array of separate
elements each radiating spherical waves. By ascertaining the points
where the waves maxima and minima meet, points of constructive and
destructive interference can be established, and the ultrasonic field
estimated. This 2-dimensional example gives some idea of the complexity
of an ultrasonic field compared with the simple “text book” shape.
The
theoretical field for a circular transducer is shown above. Moving
along the central axis of the beam away from the transducer in the near
zone, the intensity shows successive axial maxima and minima which
become further apart away from the transducer. There are also several
maxima across the beam diameter. The last axial maximum occurs at the
end of the near zone (at a distance of D2/4 lambda). Beyond this in the far
zone the central axis intensity decreases and the beam diverges.
Rectangular transducers and pulsed ultrasound complicate these fields.
It
is normal to operate ultrasound systems in the near field in order to
have a narrow beamwidth (good lateral resolution) with little
divergence. Since ultrasound can only penetrate a limited depth,
transducers are usually designed so that the end of the near field
corresponds to the limit of penetration.
Smaller
diameter crystals produce a narrower beam but at the expense of a
shorter near field and greater divergence in the far field. Higher
operating frequencies give a longer near field, but unfortunately higher
frequencies have a higher attenuation so the penetration is less.
Transducer design is therefore a compromise.
Tidak ada komentar:
Posting Komentar